You've probably seen many Top 10 lists. We have taken our own spin on the ever-popular "Top-10" list as it relates to Rigid Scope Care. Rigid scopes are particularly difficult for sterile processing technicians due to the visual component, fragility, high cost, and ease with which scopes can become damaged. Continue reading to learn more about common Rigid Scope Hazards and how employees can help prevent them!

Please visit our website for more information on this topic.

 

1. Dystal damage not seen

During a surgical procedure, the distal end of the scope may come into contact with other surgical devices such as shavers, lasers, and cautery devices, which can burn, scratch, chip, or damage the distal end glass and/or the metal outer tube and light fibers at the scope's tip.

Danger: Damaged glass can become loose or enter the patient causing poor recovery or repeat surgical procedure. Sharp edges or burs on the scope can cause tissue trauma to the patient. Fibers can become damaged or recessed, resulting in light loss and a cavity for bioburden that can make sterilization ineffective.

Prevention: Specifically check the distal end including the window, sealant/solder seal, fibers and around the entire distal end for trauma, under magnification, after every procedure and before sterilization. Any scopes with distal damage should be removed from service and repaired.

2. Poor cleaning of the dystal glass

The dystal glass is subject to bioburden, chemical surfactants, and hard water deposits that can leave residue or build up on the lens. Inadequate cleaning coupled with subsequent use and sterilization can cause the build up to become worse over time. 

Hazard: Residue on the distal window will create poor optical quality and cross-contamination risks. (Remember, if it is not clean, it cannot be sterile.)

Prevention: Clean and polish the distal window on a daily basis. You can use a manufacture provided polishing paste (be careful to rinse and remove), a water-based cleaning agent, or non-abrasive cloth after cleaning and before sterilization. Our favorite tool is a simple pencil eraser. Carefully &#;erase&#; the window to remove stubborn build up followed by an Acetone or Alcohol wipe. Check under magnification. Repeat if necessary. (You can also clean the eyepiece window using this same technique.)

3. Poor inspection of the image

A visual inspection is the most important inspection step to be sure the image is clear and ready for the next surgical procedure. Unfortunately, many technicians have not been properly trained on how to look through the scope to determine if the image is clear.

Danger: Improper inspection of the image leads to faulty scopes making it back into the surgical rotation causing decreased optical quality and increased complaints from the surgical team.

 

Prevention: After you&#;ve checked and cleaned the dystal window, look through the scope. Look down at a white, non-glossy piece of paper with fine print to be sure the entire image is crisp and clear and completely round. Many technicians simply look into space. This does not enable a view of fine image detail, nor simulate the focal distance of the anatomy being viewed during the surgical procedure. When in doubt hook the scope up to a video camera and check the image on a video monitor.

Or make it safe and use LightControl.

4. Wrong transport and storage

It is all too common for scopes to be transported to sterile processing with other instruments placed on top of the scope, or transported with the scope unsecured and rolling around inside the tray.

Danger: Increased risk of damage to the equipment and increased repair costs.

Prevention: Separate scopes from other instrumentation during transport and storage, being sure the scope is secured and unable to roll around or fall..

5. Neglecting inspection of internal optics

There is a common overlooked inspection step; the inspection of the internal optical components. When optics inside the scope become chipped, broken, or moisture filled, they can cause visualization problems ranging from a hazy/foggy image to a partial or completely black image.

Danger: Any type of optical damage can degrade the quality of the surgical procedure and increase end-user complaints. Optic damage may also coincide with other scope damage such as a misaligned insertion tube or seal failure that can cause additional problems with usage or safety.

Prevention: Using a magnification loupe or a rigid internal optical scanner such as EndoScan you can look &#;inside&#; the optics to examine the lens system and identify defects such as chipped or broken lenses, lens separation, fluid invasion and other defects. Without such a tool it you would not be able to examine the internal lenses. Additionally, it is good for the SPD department to have access to a video monitor or even a wireless camera to duplicate the image on monitor for additional checking.

6. Flash autoclaving

Due to lack of inventory, or because they have been told it is acceptable, many facilities will use Immediate Use Steam Sterilization (IUSS) in order to expedite turnover.

Danger: Flash Autoclaving of scopes on a regular basis (even if approved by the manufacturer) will degrade the life-expectancy of the scope and increase the incidence of scope repairs. The rapid heating and cooling stresses the glass optical components and hermetic seals.

Prevention: Maintain adequate inventory so that flash and routine autoclaving is not required. Using a low-temp sterilization method, although not required, can greatly extend the life-expectancy of your equipment.

7. Pressure on the scope shaft

The glass rod lenses inside the shaft of the scope are very delicate. Any pushing or pulling of the shaft/insertion tube when placing or removing scopes from the sterilization container can crack the lenses inside or bend the scope shaft. Likewise, any pressure, dings or dropping of the scope can also damage the delicate optics.

Danger: Bending the scope in any way can cause costly damage. Also a misaligned shaft can cause image shift or problems with mating instrumentation during the surgical procedure

Prevention: Do not handle the scope by the shaft. Be very careful when placing the scope in the sterilization tray. Avoid pushing or pulling on the scope shaft.

8. Lack of inventory

Have caseloads increased, or new surgeons joined the staff, but additional inventory has not been procured? This is just one indicator that you may not have adequate inventory. Other signs include; always needing loaners, delayed or cancelled cases due to insufficient equipment on hand, frequently having multiple scopes in for repair at a time, or needing to routinely use IUSS (immediate use steam sterilization) in order to turn over equipment quickly to keep up with caseload demands.    

Danger: Cancelled or delayed cases and more frequent repairs due to IUSS are common risks. Additionally, rushed staff may cut corners in the inspection and reprocessing of equipment in order to turn over equipment for the next case. This can result in increased risks to patient care and increased stress on the staff and complaints from surgical team.

Prevention: Regularly analyze and assess if you are able to keep up with growing cases or increases in demand. Fortunately, you can often purchase additional scope and instrument inventory at a fraction of the cost of new by utilizing a reputable pre-owned equipment vendor if new equipment isn&#;t in the budget. If excessive repairs are an issue, in-services and re-addressing proper care and handling with the staff can often result in a decrease in repairs. 

9. No checking of the Light Fibres

The light fibers that carry light through the scope are separate from the internal optical lenses. When the scope is &#;dark&#; you must be able to determine if the darkness is from poor optics, or damaged light fibers, as either can be the cause of diminished light.

Danger: Poor Light output from broken or damaged light fibers degrade the optical quality of the procedure. Additionally, this may cause the surgical staff to increase the light output on light source which creates additional heat and an increased risk of burns or fire from hot light cords that may touch the patient or a surgical drape.

Prevention: Visually Inspect the light fibers in the scope (and light cords) by holding the light post of the scope up to a standard light bulb and looking at the opposite (distal) end. Excessive broken fibers will be noted by black dots. Be sure to check in both directions also looking for burns, recessed or exposed fibers. Visual inspections of light are pretty subjective so we recommend measuring light efficiency using an EndoLume light meter to calculate light efficiency in the light source, light cord and scopes. 

10. Never removing light posts adaptors for Cleaning and Sterilisation


All devices with removable components should be disassembled prior to sterilization. Light post adaptors are no exception. Additionally, any operative scope with an internal channel should be cleaned as instructed. 

Danger: Corrosion and bioburden underneath the light post adaptors, or from an inadequately cleaned channel, can pose a cross contamination risk.

Prevention: Remove and place the light post adaptors or other removable accessories in the sterilization container with the scope during sterilization. Have adequate inventory on hand in case adaptors are misplaced. Lastly, never glue on the light post adaptors nor try too hard to remove adaptors that will not come off. If you twist the light post, you can sever the light fibers underneath. Follow the manufacturer IFU for instructions on cleaning scopes with an internal operating channel.

 

This blog post was originally found on LinkedIn and was written by Lee Ann Purtell  at Capital Medical Resources LLC.

 

 

If you are looking for more details, kindly visit Lingchuang Yihui.

 

What to Look For When Choosing a Flexible Endoscope

British Small Animal Veterinary Congress

Philip Lhermette, BSc(Hons), CBiol, MIBiol, BVetMed, MRCVS

Elands Veterinary Clinic

Dunton Green, Sevenoaks, Kent

 

Flexible endoscopy has been used in human and veterinary medicine for many years. Over time instrumentation has become more sophisticated and prices have reduced, encouraging more widespread use. With a vast array of new and secondhand endoscopes on the market, it is easy to make a wrong decision and be left with an expensive white elephant. No single endoscope will be suitable for all procedures in small animal practice. It is therefore important to understand your clinical requirements first of all and then to understand enough about endoscope design and function to find an instrument to meet them.

Types of Flexible Endoscope

There are two main types of flexible endoscope in common clinical use. Gastroscopes and bronchoscopes. Gastroscopes should have four-way tip deflection (at least 180 degrees in one direction), automatic air/water insufflation, a biopsy/suction channel and a working length of at least 100 cm. For large breeds a working length of 130-150 cm is essential to examine the small intestine and caecum. The diameter of the insertion tube should be no larger than 10 mm or it will not be possible to pass through the pylorus of smaller patients. A diameter of less than 8 mm makes passage through the pylorus easier in small dogs and cats but must be offset against the smaller biopsy channel--usually around 2 mm as against 2.8 mm in the larger endoscope. Smaller biopsy samples can be more difficult for pathologists to interpret.

Bronchoscopes should have two-way tip deflection of at least 100 degrees in each direction, and preferably 180 degrees in one direction to allow retroflexion over the soft palate. You will also require suction and air/water insufflation, a biopsy/working channel and a working length of at least 80 cm. An outer diameter of 2.8-5 mm is ideal for all adult dogs and cats, although a 7-8 mm gastroscope can double as a bronchoscope for medium to large dogs.

Flexible endoscopes can be further divided into fiberscope and videoendoscopes. Fibreoptic endoscopes transmit the image to the eyepiece via a bundle of coherent fibres, resulting in a pixelated image that is viewed directly through the eyepiece or via an attached camera, on a monitor. A videoendoscope has no eyepiece, the image being transmitted electronically from a CCD or CMOS video chip at the tip of the insertion tube directly to a video monitor. This gives a vastly superior image but at greater cost, since each endoscope essentially incorporates its own camera system.

Decide What Sort of Endoscope You Need

An endoscope must be fit for purpose--what is your caseload? Do you deal primarily with cats, small dogs, or large dogs? Do you see mainly gastrointestinal (GI) or respiratory cases? What quality, especially of optics, do you require? Do you feel that a warranty is important and what about servicing costs and loan of an instrument during repairs? Look at the individual characteristics of an instrument--does it feel good in your hands; is it comfortable to use?

One endoscope is rarely sufficient and you will eventually purchase another, or you may already have a range of rigid endoscopes. How can the new endoscope be integrated with current equipment and future expansion plans? Can you use the same light source and camera system on all your endoscopes to spread the cost? What is the total cost--both initial and ongoing maintenance? How are you going to recoup this?

New or Secondhand

An important decision to make is whether to buy new or secondhand. There are a lot of excellent 'veterinary specific' endoscopes appearing on the market. These are more suited to our patients than many human models, having a 130-150 cm insertion tube as standard. In addition, buying new provides you with a warranty for at least a year, and servicing and spares should not be a problem. Suppliers will usually be happy to provide training in use, care and maintenance free of charge for yourself and your staff. Proper care is vital if your investment is not to be damaged, incurring expensive repairs

Large numbers of secondhand endoscopes are available from the human market. However--'caveat emptor'--there are a number of things to look out for:

 Insertion tube length. Most human paediatric endoscopes are 100 cm, too short for GI work in large dogs.

 Insertion tube diameter >10 mm is not suitable for small animal work.

 Size of instrument channel--aim for 2.8 mm with a minimum of 2 mm. Pass an instrument through the channel to ensure there is no blockage or damage to the channel.

 Check the rubber at flexible tip for perishing or damage.

 Some older hospital endoscopes hitting the market are not fully immersible, making cleaning and adequate disinfection extremely difficult.

 Fibre damage. Remember there are light guide fibres in a video endoscope as well. Look for adequate light transmission by viewing the image with the endoscope tip placed in a large dark box. (Your hand will reflect too much light and show a good image in any case.) Check fibreoptic endoscopes for black spots on the image in the eyepiece and on a monitor, depicting broken fibres.

 Damage to guide wires. These are the wires that move the flexible tip of the endoscope. The tip should move through the whole range of movement expected--and at least 180 degrees in one direction. The tip should start moving as soon as the wheel or lever is moved, with no delay. On gastroscopes, the tip should be straight with the control wheels in the neutral position (usually with the U & L markers uppermost).

 Is the instrument water tight? This is vital as the delicate fibres and guide wires are very easily damaged by the slightest leak, and repair can cost almost as much as a new endoscope. Always attach a leak tester to the pressure compensation port and inflate to the prescribed pressure. Look for a pressure drop. If you are unsure, immerse the pressurised endoscope in water and look for escaping bubbles.

 Accessories, spares and servicing. Does the endoscope come with necessary accessories such as a leak tester, pressure compensation valve, cleaning brushes and all the necessary buttons and grommets for the working channel? Is there any documentation giving the cleaning instructions? Where will you get it serviced and are spare parts still available?

Where to Buy

New

Major manufacturers in both the human and veterinary market produce their own endoscopes. These are usually the most expensive, but are high quality instruments backed up by a large company that is unlikely to disappear, good supplies of spare parts and excellent service contracts that will often loan a replacement instrument if yours is not functional. Many veterinary resellers also offer excellent deals on instruments manufactured by smaller companies. Often the quality is not quite so high, but this is reflected in the price, and you will still get a warranty and excellent service deals.

Secondhand

Secondhand equipment can be bought direct from hospitals, via hospital auctions or even on eBay. However, purchasing flexible endoscopes this way is fraught with danger and you can easily end up with an expensive piece of junk that costs more to repair than it is worth. A number of resellers offer secondhand reconditioned equipment that will at least have been tested and will come with a short warranty. Try and persuade them to allow you to use it on trial for a month or so on real cases before purchase.

Care and Maintenance

With proper care and maintenance your endoscope should last years:

 Always clean your endoscope thoroughly immediately after use with an enzymatic cleaner. Dried debris can block channels and be extremely difficult to remove.

 Resterilise your endoscope after cleaning using gas sterilisation or a recommended cold steriliser.

 Always use a manufacturer-recommended enzymatic cleaner and cold steriliser. Inappropriate chemicals can damage seals.

 Storage. Always store your endoscope hanging with the insertion tube vertical and the buttons removed to allow the channels to drain. Fibres retain a memory if coiled and will be more prone to breakage if constantly coiled and straightened. Never store your endoscope in its case. This not only coils the insertion tube, it provides an ideal environment for bacterial growth within the endoscope. Pseudomonas is commonly isolated from incorrectly stored endoscopes.

References

1.  Chamness C. Instrumentation. In: Lhermette, P; Sobel, D. eds. BSAVA manual of endoscopy in the dog and cat. In production expected late /.

2.  Chamness C. Introduction to veterinary endoscopy and endoscopic instrumentation. In: McCarthy, T. ed. Veterinary endoscopy for the small animal practitioner. Elsevier Saunders, ; 1-20.

3.  Chamness C. Endoscopic instrumentation. In: Tams, T. ed. Small animal endoscopy, second edition. St Louis: Mosby, ; 1-16.

If you want to learn more, please visit our website types of rigid endoscopes.